Olympus Revises Duodenoscope Reprocessing Instructions

June 1, 2015

A revised reprocessing manual has been released by Olympus for the TJF-Q180V duodenoscope.

Olympus released its revised manual for reprocessing the TJF-Q180V duodenoscope, the model recently linked to several patient-to-patient infections of Carbapenem-resistant Enterobacteriaceae (CRE) bacteria associated with ERCP procedures. Information on cleaning, disinfection and sterilization of the TJF-Q180V from the Olympus website states:

"Olympus recommends you pay careful attention to the Supplemental Instructions to the TJF-Q180V Reprocessing Manual, including precleaning and alcohol flushing steps, when using automated endoscope reprocessors (AERs). Please note the setting of the elevator to the intermediate position (half up/half down) when placed in an AER. Olympus has provided major AER manufacturers with these Supplemental Instructions, and asks that you please contact the manufacturer of your AER for instructions and information relating to use of their equipment." 

Olympus has also posted an instructional video on how to reprocess its ultrasound endoscopes, which have also been linked to patient-to-patient infections following endoscopic ultrasound procedures.

Read about what AGA is doing to help our members.

More on ERCP

Attend the AGA Postgraduate Course for Practical, Take Home Information

Feb. 12, 2018

Join us on June 2-3 in Washington, D.C. This course is held in conjunction with Digestive Disease Week® (DDW) 2018.

Voluntary Recall of Pentax ED-3490TK Duodenoscopes

Feb. 8, 2018

Recall designed to reduce the potential for leakage of patient fluids into the closed elevator channel and under the distal cap.

Reprocessing of Duodenoscopes: How Do We Get to Zero?

Jan. 31, 2018

Prevention of endoscope transmitted infection will be the focus of a session at the 2018 AGA Tech Summit, a meeting for stakeholders in GI innovation.